Transmission Efficiency of the Recoil Mass Spectrometer EMMA at TRIUMF
Publication Date
5-1-2026
Document Type
Article
Publication Title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume
1085
DOI
10.1016/j.nima.2026.171295
Abstract
The mean transmission efficiency of the EMMA recoil mass spectrometer at TRIUMF has been measured with 6 different angular apertures at 17 kinetic energy/charge deviations with respect to the central, reference trajectory. Measurements performed using a 148Gd a source installed at the target position of the spectrometer are compared to ion-optical calculations and Monte Carlo simulations. The transmission efficiency as a function of angle and kinetic energy/charge is described empirically using piecewise Gaussian functions whose parameters are fit to the data.
Funding Sponsor
National Research Council of Canada; UK Science and Technologies Facilities Council (STFC)
Keywords
Recoil mass spectrometer; Transmission efficiency; ISAC-II
Department
Chemistry
Recommended Citation
B. Davids, N. E. Esker, J. Jaeyoung, Y. K. Kim, K. Pak, and M. Williams. "Transmission Efficiency of the Recoil Mass Spectrometer EMMA at TRIUMF" NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT (2026). https://doi.org/10.1016/j.nima.2026.171295