Transmission Efficiency of the Recoil Mass Spectrometer EMMA at TRIUMF

Publication Date

5-1-2026

Document Type

Article

Publication Title

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT

Volume

1085

DOI

10.1016/j.nima.2026.171295

Abstract

The mean transmission efficiency of the EMMA recoil mass spectrometer at TRIUMF has been measured with 6 different angular apertures at 17 kinetic energy/charge deviations with respect to the central, reference trajectory. Measurements performed using a 148Gd a source installed at the target position of the spectrometer are compared to ion-optical calculations and Monte Carlo simulations. The transmission efficiency as a function of angle and kinetic energy/charge is described empirically using piecewise Gaussian functions whose parameters are fit to the data.